- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
Martindale DT173 K Type Digital Thermometer
Includes: K-Type bead thermocouple, 4 x 1.5V, AAA alkaline batteries & manual.
The Martindale DT173 is a portable, single input, compact thermometer for use with Type K thermocouples.
Readings can be displayed in either Centigrade of Fahrenheit and the thermometer has an adjustable offset to optimise thermocouple performance over the desired temperature range. There is a MIN/MAX/AVG mode for recording extremes of readings. In this mode the auto power off function is disabled to allow recording over longer periods of time.
NB: The maximum temperature measurement of the DT173 is 1372°C (1999°F), but the thermocouple probe being used may only be specified to 260°C (500°F).
FEATURES
Temperature range -50°C to 1372°C (-58°F to 1999°F).
Thermocouple compensation adjustment.
Max / Min function.
Relative function.
Data hold function.
Backlight.
Auto power off.
Electrical | |
All specified accuracies are at 23°C ± 5°C, <75% RH for 1 year, not including thermocouple error. | |
Temperature coefficient | 0.1 x (applicable accuracy) per °C. (0°C to 18°C, 28°C to 50°C). |
Temperature scale | Celsius or Fahrenheit, user-selectable. |
Thermocouple type | K. |
Measurement range | -200°C to 1372°C. (-328°F to 1999°F). |
Note: Measurement range will be limited by the range of the thermocouple probe used. | |
Resolution | 0.1° to 199.9°C/°F, 1° t 200°C/°F. |
Temperature Accuracy | |
-200°C to -60°C | ± (0.1% of rdg + 2°C). |
-60°C to 1372°C | ± (0.1% of rdg + 1°C). |
-328°F to -76°F | ± (0.1% of rdg + 4°F). |
-76°F to 1999°F | ± (0.1% of rdg + 2°F). |
Input protection | 24V DC or AC rms maximum input voltage on any combination of input pins. |
Input connector | Accepts standard miniature thermocouple connectors (flat blades spaced 7.9mm. centre to centre). |
Supplied thermocouple probe | 4-foot Type K thermocouple bead probe (Teflon tape insulated). |
Maximum insulation temperature | 260°C (500°F). |
Probe accuracy | ± 2.2°C or ± 0.75% of reading (whichever is greater). |
Environmental | |
Temperature & Humidity | (Operating): 0°C to 50°C, <70% R.H. (Storage): -20°C to 60°C, 0 to 80% R.H., batteries removed. |
Altitude | up to 2000m. |
Pollution degree | 2. |
Safety | Conforms to BS EN 61010-1. |
Class II, double insulation. | |
EMC | Conforms to BS EN 61326-1. |
General Specifications |
|
Display | 3½ digit liquid crystal display (LCD) with a maximum reading of 1999. |
Measurement rate | 2.5 times/second. |
Polarity | Automatic, positive implied, ‘-‘ for negative polarity indication. |
Overrange | (OL) is displayed. |
Power | 4 x 1.5V, AAA alkaline batteries (IEC LR03, NEDA 24A). |
Battery life | 200 hours typical with alkaline batteries. |
Low battery indication | Low battery symbol is displayed. |
Auto power off | After 30 minutes. |
Dimensions | 160 mm (H) x 83 mm (W) x 38 mm (D). |
Weight | Approx. 240g, including batteries. |
Martindale DT173 warranty | 2 years. |