- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
Fluke 1551A-12 Intrinsically Safe 'Stik' Thermometer
Includes: NVLAP-accredited report of calibration, Fluke 1551A User’s Guide on CD-ROM, 3 AAA batteries.
Finally, a digital substitute for your mercury-in-glass thermometers! Accurate and repeatable to ± 0.05 °C over its full range, the Fluke 1551A 'Stik' Thermometer is the new 'gold standard' of industrial temperature calibration. Whether working outdoors in environments where potentially explosive gases may be present or on the floor of a processing plant, the intrinsically-safe, battery operated, portable reference thermometer is designed to go where you work.
The large backlit LCD display on the Fluke 1551A Ex rotates 90° degrees making it easy to read from any angle. A user-configurable stability/trend indicator lets you know when your temperature is stable enough to record an accurate measurement. A user-settable auto-off function extends typical battery life to three hundred hours. A low-battery indicator and stop-measure function prevent erroneous measurements from being made due to low battery life.
The sensor of the Fluke 1551A Ex requires a minimum immersion depth of only seven centimeters (2.8 inches), with no appreciable effect on the measured temperature due to heat loss through stem conduction.
FEATURES
Accuracy of ± 0.05 °C (± 0.09 °F) over full range.
Intrinsically safe (ATEX and IECEx compliant).
Two models to choose from (-50 °C to 160 °C or -80 °C to 300 °C).
User-configurable temperature trend/stability indicator.
Display temperature in °C or °F.
Optional data logging to internal memory.
300-hour battery life.
Percent battery-life and low-battery indicator.
NVLAP-accredited, NIST-traceable calibration included.
FIVE REASONS TO REPLACE MERCURY THERMOMETERS
1. Since the “Stik” Thermometer does not contain mercury, it is not subject to government-imposed bans. Government agencies worldwide have either banned or are seeking to ban the use and transportation of industrial mercury thermometers.
2. The stainless steel probe sheath of the “Stik” Thermometer is far more robust than the glass stem of a mercury thermometer and better suited for industrial environments.
3. When considering cost of ownership, a digital thermometer is less expensive than a mercury thermometer. With a mercury thermometer, you either invest in equipment or you hire out to clean up after a mercury spill.
4. Mercury thermometers cannot be adjusted after calibration. The actual temperature must be calculated by applying corrections to the measured temperature.
5. A digital thermometer doesn’t require that you match calibration conditions to achieve an accurate measurement. You’re only required to meet the minimum immersion requirements - which is only 7 cm (2.8 in) in the case of the 1551A Ex.
Temperature Range | -50 °C to 160 °C (-58 °F to 320 °F). | ||
Accuracy (1 year) | ± 0.05 °C (± 0.09 °F). | ||
Display units | °C, °F. | ||
Sensor type | 100 Ω thin-film RTD. | ||
Probe temperature coefficient | 0.00385 Ω/Ω/°C nominal. | ||
Sensor length | ≤ 10 mm. | ||
Sensor position (from sheath tip) | 3 mm (0.1 in). | ||
Minimum immersion depth | 7 cm (2.8 in). | ||
Probe sheath material | Stainless steel. | ||
Response time | 4.8 mm diameter probe (3/16 in): 14 seconds 6.35 mm diameter probe (1/4 in): 21 seconds. |
||
Probe hysteresis | ± 0.01 °C. | ||
Temperature resolution | Selectable: 0.1, 0.01, 0.001 (default 0.01). | ||
Sample rate | Selectable: 0.5 seconds, 1 seconds, 2 seconds (default 1 seconds). | ||
Operating temperature range readout | -10 °C to 50 °C (14 °F to 122 °F). | ||
Storage temperature | -20 °C to 60 °C (-4 °F to 140 °F) 0 % to 95 % RH, non-condensing. | ||
Damping | Moving average of the most recent 2 to 10 readings (ON/OFF, 2, 5, 10). | ||
Communications | RS-232 stereo jack (access calibration parameters only). | ||
AC power | None. | ||
DC Power | 3-AAA Batteries, typical battery life of 300 hours without LCD backlight. | ||
EMC compliance | EN61326:2006 Annex C; CISPR II Edition 5.0-2009; Class “B. | ||
Enclosure rating | IP50. | ||
Electronics dimensions (H x W x D) | 114 mm x 57 mm x 25 mm (4.5 in x 2.25 in x 1.0 in). | ||
Weight | 196 g (6.9 oz). | ||
Calibration (included) | NVLAP-accredited, NIST-traceable. | ||
Characterisation | CVD. |
No customer reviews for the moment.
COMPONENT |
DESCRIPTION | PART No. | ||
Probe & Clip Set | 10 A Fused test probe and clip lead set. | 1013-224 | ||
Lead Set | DLRO2 current and potential leadset 2m. 2 x red lead, 2 x blk lead, 2 x grabber clip, 2 x probe. | 1011-673 | ||
Right-angled Adaptors | 4 right angled adaptors to allow hook terminated (E.g. KC100) leads to fit DLRO2. | 1012-511 | ||
Probe Pins/Tips | Set of 4 Kelvin probe pins. Replacement probe tips. | 1012-064 | ||
50m Cable Reel | KC50C-KIT Kelvin clamp/probe reel (50 m). | 1013-211 | ||
100m Cable Reel | KC100C-KIT Kelvin clamp/probe reel (100 m). | 1013-212 | ||
50m Ext. Reel | KC50E Extension reel (50 m). | 1013-213 |