Fluke 725Ex Intrinsically Safe Process Calibrator
  • Fluke 725Ex Intrinsically Safe Process Calibrator
  • Fluke 725Ex Intrinsically Safe Process Calibrator
  • Fluke 725Ex Intrinsically Safe Process Calibrator

Fluke 725Ex Intrinsically Safe Process Calibrator

Fluke authorised distributor and calibration house.

Includes: AC72 alligator clips, one set of stackable test leads, TL75 testleads, Fluke 725 Ex User manuals and NIST- traceable calibration report and data.

2098179
Call For Price

The Fluke 725Ex is obsolete.

The Fluke 725Ex Calibrator has been designed for use in Ex Hazardous Areas. These are areas where potentially flammable or explosive vapors may occur. These areas are referred to as hazardous (classified) locations in the United States, as Hazardous Locations in Canada, as Potentially Explosive Atmospheres in Europe and as Explosive Gas Atmospheres by most of the rest of the world.


The Model 725 Ex calibrator is designed as intrinsically safe. This means that connecting the Fluke 725Ex calibrator to equipment that is used within intrinsically safe circuits will not cause an ignition capable arc as long as the entity parameters are suitably matched.

The Fluke 725Ex Intrinsically Safe Multi-function Process Calibrator is powerful yet easy-to-use. Measure and source mA, volts, temperature (RTDs and thermocouples), frequency and ohms. Combined with one of the (8) Fluke-700PEx pressure modules, the 725Ex is able to calibrate almost any process instrument likely to need service in an area where explosive gasses may be present.

Do you need to calibrate transmitters? The Fluke 725Ex split display lets you view input and output values simultaneously. For valve and I/P tests, you can source mA while measuring pressure. The 725 Ex has auto-stepping and auto-ramping for remote testing, plus 25% stepping for fast linearity tests.

FEATURES

ATEX II 1 G Ex ia IIB 171°C KEMA 04ATEX 1303X.
I.S. Class I, Division 1 Groups B-D, 171°C compliance.
Measure Volts DC, mA, RTDs, thermocouples, frequency and ohms.
Source or simulate Volts dc, mA, RTDs, thermocouples, frequency and ohms.
Two channel simultaneous source and measure capability for calibration of transmitters.
Power transmitters with internal loop supply.
Store frequently-used test setups for later use.
Pressure measurement to 3,000 psi/200 bar using any of the 8 intrinsically safe Fluke 700PEx Pressure Modules.
Pressure switch test function to capture set, reset and deadband values.
Compact size and weight.
Simple, push-button user interface.
Rugged and reliable, for field use.

Fluke 725 Ex Specifications
Measurement Accuracy
Voltage DC 30.000 V 0.02% + 2 counts (upper display).
10.000 V 0.02% + 2 counts (lower display).
90.00 mV 0.02% + 2 counts.
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector).
Current DC 24.000 mA 0.02% + 2 counts.
Resistance 0.0 to 400.0 Ω 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire).
401 to 1500 Ω 0.5 Ω (4-wire), 1 Ω (2- and 3-wire).
1500 to 3200 Ω 1 Ω (4-wire), 1.5 Ω (2- and 3-wire).
Frequency 2.0 to 1000.0 CPM 0.05% + 1 count.
1.0 to 1100.0 Hz 0.05% + 1 count.
1.00 to 10.00 kHz 0.05% + 1 count.
Sensitivity 1 V peak-to-peak-minimum.
Pressure Accuracy from 0.025% of range using any of 8 intrinsic safe pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC 100.00 mV 0.02% + 2 counts.
10.000 V 0.02% + 2 counts.
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector).
Current DC 24.000 mA (source) 0.02% + 2 counts.
24.000 mA (simulate) 0.02% + 2 counts.
Resistance 15.0 to 400.0 Ω 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA).
401 to 1500 Ω 0.5 Ω (excitation current 0.05 to 0.8 mA).
1500 to 3200 Ω 1 Ω (excitation current 0.05 to 0.4 mA).
Frequency 2.0 to 1000.0 CPM 0.05%.
1.0 to 1100.0 Hz 0.05%.
1.00 to 10.00 kHz 0.25%.
Waveform 5 V p-p squarewave, -0.1 V offset.
RTDs & Thermocouples
Measure accuracy NI-120 0.2°C.
PT-100 (385) 0.33°C.
PT-100 (393) 0.3°C.
PT-100 (JIS) 0.3°C.
PT-200 (385) 0.2°C.
PT-500 (385) 0.3°C.
PT-1000 (385) 0.2°C.
Resolution 0.1°C.
J 0.7°C.
K 0.8°C.
T 0.8°C.
E 0.7°C.
R 1.8°C.
S 1.5°C.
B 1.4°C.
L 0.7°C.
U 0.75°C.
N 0.9°C.
Resolution J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F.
XK 0.6°C.
BP 1.2°C.
Source accuracy NI-120 0.2°C.
PT-100 (385) 0.33°C.
PT-100 (393) 0.3°C.
PT-100 (JIS) 0.3°C.
PT-200 (385) 0.2°C.
PT-500 (385) 0.3°C.
PT-1000 (385) 0.2°C.
Resolution 0.1°C.
Accuracy stated for 4-wire measurement.
J 0.7°C.
K 0.8°C.
T 0.8°C.
E 0.7°C.
R 1.4°C.
S 1.5°C.
B 1.4°C.
L 0.7°C.
U 0.75°C.
N 0.9°C.
Resolution J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C.
XK 0.6°C.
BP 1.2°C.
Specifications
Ramp functions Source functions Voltage, current, resistance, frequency, temperature.
Ramps Slow ramp, Fast ramp, 25% step-ramp.
Loop power function Voltage 12 V.
Accuracy 10%.
Maximum current 22 mA, short circuit protected.
Step functions Source functions Voltage, current, resistance, frequency, temperature.
Steps 25% of range, 100% of range.
Safety Specifications
Agency approvals ATEX II 1 G Ex ia IIB 171°C.
I.S. Class I, Division 1 Groups B-D.
Mechanical & General Specifications
Size 130 x 236 x 61 mm (5.188 x 9.291 x 2.402 in).
Weight 0.85 kg (1.874 lbs.).
Batteries 4 AA alkaline batteries.
Fluke 725Ex Warranty Three year.
Battery replacement Separate battery compartment, accessible without breaking calibration seal.
Side port connections Pressure module connector.

Download

Fluke 725Ex Datasheet

PDF Specifications of the Fluke 725Ex Intrinsically Safe Process Calibrator.

Download (254.87k)