Time 1090 Portable Process Calibrator
Includes: Time 1090, carry case, rechargeable battery pack (installed) with mains charger, and user manual.
The Time 1090 is a high performance instrument designed for use in both the lab and field applications. It simulates and can measure from the most commonly used temperature sensors - thermocouples (8 types) and PT100. In addition it can generate and measure mV and mA.
Thermocouple Measurement and Simulation
Measure and simulate the temperature and mV characteristics of J, K, T, R, S, B, N and E thermocouples.
Cold Junction Compensation
The unit can be operated with or without internal cold junction compensation.
PT100 Measurement and Simulation
Based on 0.3850 alpha probe standard. Range is -200°C to 700°C.
Measurement and Source (µV, mV, and mA)
Measurement ranges are 0 to +/-30mV and 0 to +/-60mA. Source ranges are 0 to +/-80mV and 0 to +80mA.
Temperature Units Selection
The display can be easily changed from °C to °F. The equivalent uV (thermocouples) and ohms (PT100) can also be shown.
24V Process Loop Drive Mode
A process loop can be driven at 24V and up to 60mA by selecting the ‘Milliamp Source’ mode and setting it at 60mA (or a lower level if required).
Inching (Incrementing/Decrementing)
The unit has a general-purpose inching function. This adjusts the output in fixed increments of temperature (thermocouples only) or voltage or current. The set-up menu gives a the user a choice of three levels of increment i.e. 0.1, 1 or 10 for °C/°F, or 1, 10, or 100 uV/uA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output. This is especially useful for calibrating thermostat controllers that have tight specification on hysteresis.
Memory Recall and Step/Auto-Step Functions
Up to 10 values can be stored in the unit’s non-volatile memory and they can be recalled at any time. The user can also manually step through them in sequence using the step key. Continuous stepping (auto-step) is also available at any user selectable rate between 1 & 10 seconds/step.
Power is via an internal high capacity rechargeable metal hydride battery that can be recharged from the supplied external mains charger. An auto power-down feature helps conserve battery life by switching off the instrument if inactive for over 5 minutes. This feature can be disabled if not required.
FEATURES
Measure/simulate thermocouples.
Measure/simulate PT100.
Measure/source (µV/mV/mA).
Display in °C and °F.
Automatic CJC - selectable.
10 point memory recall.
Process loops 4-20mA/0-50mA.
Temperature Measure | |||
---|---|---|---|
Thermocouple Type | Temperature Range (°C) | Accuracy (°C) | |
J | -200 to 580 | 0.7 | |
K | -200 to -150 -150 to 750 |
2.5 0.5 |
|
T | -200 to 0 0 to 400 |
1.5 0.4 |
|
R | 50 to 400 400 to 1750 |
3.0 1.5 |
|
S | -50 to 100 100 to 1750 |
3.0 1.5 |
|
B | 110 to 1000 1000 to 1800 |
3.5 1.5 |
|
N | -100 to 890 | 0.6 | |
E | -50 to 400 | 0.4 | |
Resolution | 0.1°C or 0.1°F | ||
Temperature Simulation | |||
Thermocouple Type | Temperature Range (°C) | Accuracy (°C) | |
J | -210 to 150 150 to 1200 |
0.15 0.3 |
|
K | -270 to 190 190 to 1250 |
0.5 0.4 |
|
T | -200 to 150 150 to 400 |
0.4 0.5 |
|
R | -50 to 800 800 to 1750 |
0.8 2.0 |
|
S | -50 to 850 850 to 1750 |
0.9 2.0 |
|
B | 100 to 1200 1200 to 1800 |
2.0 3.0 |
|
N | -270 to 260 260 to 1300 |
0.5 1.0 |
|
E | -50 to 1000 | 0.3 | |
Resolution | 0.1°C or 0.1°F | ||
Note: An additional correction representing the equivalent 1μV should be allowed for stray thermal emf effects (Temperature measure & simulation). | |||
PT100 Simulation | |||
14 set temperature points (°C) | Accuracy | ||
-100, -50, -20, 0, 20, 50, 100, 200, 300, 400, 500, 600, 700, 800 | 0.1% of resistance value (typically 0.5 °C) | ||
PT100 Measure | |||
Range | Accuracy | Resolution | |
200 to 700 °C, 2 wire. | 0.2% of resistance value (typically 0.7 °C) | 0.2°C (or 0.2°F) | |
Millivolt Measure | |||
Range | Resolution | Accuracy | Input Resistance |
0 to +/- 30 mV | 10 μV | 0.05% of full scale ±1 digit | 100 kΩ |
Millivolt Source | |||
Range | Resolution | Accuracy | Output Resistance |
0 to 8 mV | 0.5 μV | +/- 4 μV | 10 Ω |
8 to 80 mV | 5 μV | 0.02% of full scale | 10 Ω |
Milliamp Measure | |||
Range | Resolution | Accuracy | Input Resistance |
0 to 60 mA | 20 μA | 0.05% of full scale ±1 digit | 0.5 Ω |
Milliamp Source | |||
Range | Resolution | Accuracy | |
0 to 8 mA | 0.5 μA | +/-10 μA | |
8 to 80 mA | 5 μA | 0.02% of full scale | |
Further Features and Functions | |||
Inching | Three levels of increment, 0.1, 1 or 10 for °C/°F, and 1, 10, or 100 μV/ μA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output. | ||
24 V Process Loop Drive | A process loop can be driven at 24V and up to 60 mA by selecting the ‘Milliamp Source’ mode and setting it at 60 mA (or a lower level if required). Max Load Specifications: 300Ω/80mA, 480 Ω/50 mA, 1.2 kΩ/20 mA. |
||
Memory recall & step | 10 memory locations for non-volatile storage of values. Manual & AutoStep, rate adjustable from 1 to 10 sec/step. | ||
General Specification | |||
Cold Junction Compensation | Accuracy 0.2°C. Resolution 0.1°C. | ||
Operating Temperature | -10 to 40°C (15 to 105°F). | ||
Connections | Industry standard 4 mm screw terminals. | ||
Power | A Ni-MH rechargeable battery pack gives approx. 60 hours continuous use. The mains recharger supplied gives full recharge in 11 hours. To conserve battery life a user inactivity power-down feature is included. | ||
Dimensions | H235 x W150 x D75mm. | ||
Weight | 1.2 kg. | ||
Optional Extras | Calibration certificates traceable NPL and UKAS. | ||
Country of Origin | UK. | ||
Time 1090 warranty | One-year. |