The GE Druck UPS-II loop calibrator (Druck UPS-2) is an advanced handheld process calibrator tool. It can power (24 VDC) and read 2-wire transmitters to perform field calibrations. Output currents are adjustable to a resolution of 10 microamps in both source and transmitter simulation. The calibrator can source or simulate fixed calibration currents to test or calibrate receivers and to stroke valves. Different current levels are available from the menu.
Output currents can be changed manually or automatically. The time between steps is 10 seconds in the auto-stepping mode. In the valve mode, auto-stepping will be: 3.8, 4, 4.2 or 19, 20, 21 mA. The UPSII calibrator can also source or simulate with high resolution a continuous up/down cycle between 4 and 20 mA or 0 and 20 mA. The travel time between limits is 60 seconds.
FEATURES
Sources and reads milliamps.
Powers and reads 2-wire transmitters.
Reads and simulates DP transmitters in % of flow.
Auto-ramp cycling for endurance tests.
Auto-stepping for hands-free calibration.
Fixed calibration currents.
Fixed currents for valve stroking.
Fixed Steps
The calibrator can source or simulate fixed calibration currents to test or calibrate receivers and to stroke valves. Different current levels are available from the menu.
4 to 20 mA linear 4, 8, 12, 16, 20 mA.
0 to 20 mA linear 0, 5, 10, 15, 20 mA.
4 to 20 mA sq.rt (flow) 4, 5, 8, 13, 20 mA.
0 to 20 mA sq.rt (flow) 0, 1.25, 5, 11.25, 20 mA.
4 to 20 mA 3.8, 4, 4.2, 12, 19, 20, 21 mA.
Output currents can be changed manually or automatically. The time between steps is 10 seconds in the auto-stepping mode. In the valve mode, auto-stepping will be: 3.8, 4, 4.2 or 19, 20, 21 mA.
Ramping
The calibrator can source or simulate with high resolution a continuous up/down cycle between 4 and 20 mA or 0 and 20 mA. The travel time between limits is 60 seconds.
Reading in Percentage
The calibrator can display readings in percentage of a 4 to 20 mA or 0 to 20 mA span and can also read or simulate dP-transmitters directly in percentage of flow.
Output Range | 0 to 21.00 mA. | ||
Input Range | 0 to 21.00 mA. | ||
Accuracy | 0.05% of range. | ||
Resolution | 10 microamps. | ||
Temperature Effect | 0.0015% /°F (0.003% /°C). | ||
Internal Loop-Power | 24 VDC stabilised. | ||
External Loop-Power | 56 VDC maximum. | ||
Measurement Input | 12 Ω including fuse. | ||
Drive Capability | 900 Ω. | ||
Operating Temperature | 15°F to 120°F (-10°C to 50°C). | ||
Batteries | 4x 1.5V AA Batteries. | ||
Battery Life | Sources 12 mA: 13 hours Measures: 64 hours. |
||
Low Battery | Flashes “lobat”. | ||
Housing | High impact ABS. | ||
Dimensions | 3 in x 5 in x 0.83 in (77 mm x 124 mm x 21 mm). | ||
Weight | 9 oz (285 g) including batteries. | ||
Druck UPS-II warranty | One year. |
No customer reviews for the moment.
COMPONENT |
DESCRIPTION | PART No. | ||
Line adaptor | 230V UK PLG line adaptor. | UNO-13600-1 | ||
Line adaptor | Line Adapt/Charger (230V, 50Hz) EU. | UNO-13600-2 | ||
Line adaptor | Line Adapt/Charger (115V, 50Hz) USA. | UNO-13600-3 | ||