loading spinner
Time 1006 DC MilliVolt Source
  • Time 1006 DC MilliVolt Source
  • Time 1006 DC MilliVolt Source
  • Time 1006 DC MilliVolt Source
  • Time 1006 DC MilliVolt Source

Time 7005 Process Loop Calibrator

Time Authorised distributor.

Includes: Mains battery charger, and user manual.

7005
£749.00
£898.8 Inc. VAT
In Stock

A precision handheld instrument primarily used for the calibration and simulation of voltage and current loops. The 7005 is a high accuracy loop calibrator incorporating source and measure capabilities. The compact and lightweight design make it easily portable and ideal for plant operations and field use. Combined with user friendly controls and simple operation make this loop calibrator an excellent model for both process engineers and calibration technicians.

In source mode the range can be user programmed to any value between 0 mA and 50 mA, or 0 V and 21 V. For example a low point of 10 mA and a high point of 50 mA could be set giving a span of 40 mA.

Measure mode provides both voltage and current measuring capability with 5 digit resolution. Ranges are 0 to ± 5 V and ± 5 to ± 25 V, 0 to ± 25 mA and ± 25 to ± 125 mA. Alternatively the signal can be measured as a % of span for the following ranges; 4 to 20 mA, 0 to 20 mA, square root 4 to 20 mA, or square root 0 to 20 mA. For all measurements a Min/Max recording function is available on demand.

FEATURES

Loop source/measure - V or I.
Measure 0 to 25 V or 0 to 125 mA.
Source 0 to 21 V or 0 to 50 mA.
Accuracy 0.01 %.
Resolution 1 μA or 1 mV.
Programmable ranges.
Manual step, auto-step & ramp.

Time 7005 Process Loop Calibrator Specifications
Voltage Source
Range Resolution Accuracy Output Current Output Resistance
0 to 21V 1mV ±0.01% of setting ±4mV 50mA < 1Ω
Voltage Measure 0 to ±25V, Autoranging)
Range Resolution Accuracy Measure Load
0 to 5V 0.1mV ±0.01% of reading ±0.4mV 10MΩ
5 to 25V 1mV ±0.01% of reading ±2mV 10MΩ
Current Source
Range Resolution Accuracy Output Voltage Loop Resistance
0 to 50mA 1μA ±0.01% of setting ±2μA 22V max 1100Ω @ 20mA max
Current Measure (0 to ±125mA, Auto-Ranging)
Range Resolution Accuracy Measure Load
0 to 25mA 1μA ±0.01% of reading ±2μA 24.5Ω
25 to 125mA 10μA ±0.01% of reading ±20μA 24.5Ω
Current Sink
Range Resolution Accuracy Min External Drive Max External Drive
0 to 50mA 1μA ±0.01% of setting ±2μA 4V 40V
General Specifications
Operating Temperature -10 to 50 °C.
Storage Temperature -30 to 70 °C.
Power Internal rechargeable NiMH battery pack.
Battery life Typically 9 hours.
Dimensions H 165 x W 90 x D 45 mm.
Weight 0.42 kg.

Download

Time 7005 Datasheet

PDF Specifications of the Time 7005 Process Loop Calibrator.

Download (774.58k)