loading spinner
Clare HAL103 AC/DC Hipot Tester
  • Clare HAL103 AC/DC Hipot Tester
  • Clare HAL103 AC/DC Hipot Tester

Clare HAL 103 AC/DC Hipot Tester

Clare authorised distributor and service centre.

Includes: HAL102 Instruction manual, HT Clip Lead, Hipot Probe & Lead, Power Cord, Guard Plug, Comms cable.

H103
£4999.00
£5998.8 Inc. VAT

The Clare Hal 103 Safety Electrical Tester (formerly the HAL combi) is a powerful tool to assist in the analysis of the safety of electrical and electronic equipment.  A range of tests are provided, with innovative features to aid difficult test situations, which allow testing of a wide variety of equipment.

The HAL 103 is designed for simple manual testing and automatic mode for large volume applications in the most demanding environments. It is designed for use in manufacturing, laboratory, maintenance and high speed production environments either as a stand-alone test instrument or as an integrated part of a production line system.

The HAL 103 has DC and selectable 50 or 60 Hz outputs for flash testing and incorporates a 0-40A constant current low voltage ground bond test. Other features include a large high contrast bar graph display of the test output and measured test parameters. The Clare HAL103 is a fast and efficient automatic industrial test solution, which records results and enables accurate testing to be carried out in line with relevant standards at all times.

FEATURES

Multi-voltage tester.
Highly accurate Ground Bond, Hipot leakage and DC IR measurements.
Highly accurate leakage and DC IR measurements.
4 wire Ground Bond test measurements.
Usable with the Clare No – Burn Ground Bond probe.
Large Full graphics display, Real-time clock.
Connects to PC, Barcode reader and printer.
Compatible with Seaward PATguard and other leading software packages.
Software adjusts the user interface to differing levels of expertise.
Auto test code generation.
Uses fully Barcode reading for simple & easy tests initiation.
Fast Start-up.
Fully isolated test outputs, to comply with EN50191.

HAL SERIES COMPARISON CHART
FEATURES 101 102 103 104 LED
Earth/Ground Bond
AC Flash/Hipot
DC Flash/Hipot
DC Insulation Resistance
ARC Detection
Leakage
Load Power
Load Power (200mVA resolution)
Power Factor
Results Memory
Automation Option
Barcode scanner/printer Option
Internal scanner
Clare HAL Series Product Video

Clare HAL 103 Specifications
Earth Ground/Bond Test
Test Voltage 6 V AC (nominal).
Frequency 50 or 60 Hz (independent of supply).
Display Range and Accuracy 0 to 1500 mΩ - ±2% ±5 counts.
Display Resolution 1 mΩ.
Compliance Test Ranges
Current/Load Resistance/Time 5 A, 1000 mΩ - continuous
10 A, 500 mΩ - continuous
25 A, 200 mΩ - continuous
30 A, 150 mΩ - 60 seconds
40 A, 100 mΩ - 60 seconds.
Settable Output Current Range 0.1 to 40.0 A.
Selectable Range of PASS/FAIL Levels 0 to 1500 mΩ.
Insulation Resistance Test
DC Output Voltage 0.10 kV to 6.00 kV
(10 V/step when manually adjusted)
(50 V/step in test sequence program).
Display Range 0.01 MΩ to 500 MΩ.
Display Range/Accuracy 0.03 MΩ to 350 MΩ: ±5% ± 5 counts
350 MΩ to 500 MΩ: indication only.
Display Resolution 0.01 MΩ.
PASS/FAIL Level 0.00 MΩ to 500.0 MΩ.
AC Dielectric Strength Test
Programmable Voltage Range 0.10 to 5.00 kV
(10 Vstep when manually adjusted)
50 V/step in test sequence program).
Frequency (Independent of Supply) 50 or 60 Hz.
Voltage Display Range/Accuracy 0.10 kV to 5.00 kV: ±1% ±5 counts.
Voltage Display Resolution 0.01 kV.
Current Display Range/Accuracy 0.01 mA to 20.00 mA: ±1% ± 5 counts.
Current Display Resolution 0.01 mA.
Selectable Range of PASS/FAIL Levels 0.01 mA to 20.00 mA.
Maximum Current Output 20.00 mA @ 5 kV.
Optional Arc Detection 9 levels.
DC Dielectric Strength Test
Programmable Voltage Range 0.10 to 6.00 kV
(10 V/step when manually adjusted)
(50 V/step in test sequence program).
Voltage Display Range and Accuracy 0.10 to 6.00 kV: ±1% ±5 counts.
Voltage Display Resolution 0.01 kV.
Current Display Range and Accuracy 0.01 mA to 10.00 mA: ±1% ±5 counts.
Current Display Resolution 0.01 mA.
Selectable Range of PASS/FAIL Levels 0.01 mA to 10.00 mA.
Maximum Current Output 10.00 mA.
Optional Arc Detection 9 levels.
General Specifications
Operating Voltage Range Selectable: 115 to 230 V AC RMS, 50/60 Hz.
EN50191 Compliant when used with appropriate accessories.
Dimensions 200 x 300 x 370 mm (W x H x D).
Weight 15 kg approx.
Clare HAL 104 warranty 1 year warranty subject to product registration with Seaward.

Download

Clare Hal Datasheet

PDF Specifications of the Clare Hal Series Multifunction Safety Testers.

Download (1.78M)

No customer reviews for the moment.

Write your review

Clare HAL 103 AC/DC Hipot Tester
COMPONENT PARTS DESCRIPTION PART No.
Earth Test Probe Earth probe compatible with HAL series. 01520/1
Hipot Test - Probe Test probe for use with HAL series - 2 metres. 03914/2
Handheld Guard Switch Handheld guard switch compatible with the HAL series. DCS317
Warning Beacon Status Beacon For Hal Series. H-5017