Clare HAL 101 AC/DC Hipot Tester
Includes: HAL101 Instruction manual, HT Clip Lead, Hipot Probe & Lead, Power Cord, Guard Plug, Comms cable.
The Clare Hal 101 Tester is a powerful tool to assist in the analysis of the safety of electrical and electronic equipment. A range of tests are provided, with innovative features to aid difficult test situations, which allow testing of a wide variety of equipment.
The large display and simple push-button operation makes the process of testing straightforward and fast, ensuring your production line continues to efficiently meet the demands of any number of quality and safety standards.
FEATURESMulti-voltage tester.
Stores up to 6000 results.
Highly accurate leakage and DC IR measurements.
Large full graphics display.
Real-time clock.
Connects to PC, Barcode reader and printer.
Compatible with Safety E-Base and other leading software packages.
Auto test code generation.
Fast start-up.
Fully isolated outputs, to comply with EN50191.
HAL SERIES COMPARISON CHART
FEATURES | 101 | 102 | 103 | 104 | LED | |
Earth/Ground Bond | ||||||
AC Flash/Hipot | ||||||
DC Flash/Hipot | ||||||
DC Insulation Resistance | ||||||
ARC Detection | ||||||
Leakage | ||||||
Load Power | ||||||
Load Power (200mVA resolution) | ||||||
Power Factor | ||||||
Results Memory | ||||||
Automation Option | ||||||
Barcode scanner/printer Option | ||||||
Internal scanner |
Clare HAL Series Product Video
Clare HAL 101 Specifications:
Insulation Resistance Test | |||
DC Output Voltage | 0.10 kV - 6.00 kV (10 V/Step when manually adjusted) (50 V/Step in test sequence program). |
||
Display Range | 0.01 MΩ - 500 MΩ. | ||
Display Range / Accuracy | 0.03 MΩ - 350 MΩ ± 5% ± 5 Counts 350 MΩ - 500 MΩ Indication Only. |
||
Display Resolution | 0.01 MΩ. | ||
Pass/Fail Level | 0.00 MΩ - 500.0 MΩ. | ||
AC Dielectric Strength Test | |||
Programmable Voltage Range | 0.10 kV - 5.00 kV (10 V/Step when manually adjusted) (50 V/Step in test sequence program). |
||
Frequency Independent of Supply | 50 or 60 Hz. | ||
Voltage Display Range and Accuracy | 0.10 kV - 5.00 kV ±1% ± 5 Counts. | ||
Voltage Display Resolution | 0.01 kV. | ||
Current Display Range and Accuracy | 0.01 mA-20.00 mA ±1% ± 5 Counts. | ||
Current Display Resolution | 0.01 mA. | ||
Selectable Range of Pass/Fail Levels | 0.01 mA-20.00 mA. | ||
Maximum Current Output | 20.00 mA @5 kV. | ||
Optional Arc Detection | 9 Levels. | ||
DC Dielectric Strength Test | |||
Programmable Voltage Range | 0.10 kV - 6.00 kV (10 V/Step when manually adjusted) (50 V/Step in test sequence program). |
||
Voltage Display Range and Accuracy | 0.10 kV - 60.00 kV ±1% ± 5 Counts. | ||
Voltage Display Resolution | 0.01 kV. | ||
Current Display Range and Accuracy | 0.01 mA - 10.00 mA ±1% ± 5 Counts. | ||
Current Display Resolution | 0.01 mA. | ||
Selectable Range of Pass/Fail Levels | 0.01 mA - 10.00 mA. | ||
Maximum Current Output | 10.00 mA. | ||
Optional Arc Detection | 9 Levels. | ||
General Specifications | |||
Operating Voltage Range | Selectable 115 V - 230 V AC RMS, 50/60 Hz. | ||
Storage | Up to 6000 results including details of Products, user, location, site, date and time. | ||
Dimensions | 200(W) x 300(H) x 370(D) mm. | ||
Weight | 15 kg approx. | ||
Clare HAL 101 warranty | 12 months. |
No customer reviews for the moment.
COMPONENT PARTS | DESCRIPTION | PART No. | ||
Earth Test Probe | Earth probe compatible with HAL series. | 01520/1 | ||
Hipot Test - Probe | Test probe for use with HAL series - 2 metres. | 03914/2 | ||
Handheld Guard Switch | Handheld guard switch compatible with the HAL series. | DCS317 | ||
Warning Beacon | Status Beacon For Hal Series. | H-5017 |