- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
TPI 665 Dual Input Differential Manometer With Data Logging
Includes: TPI 665 manometer, Docking Station, S/Ware, 2x 1.5 Volt Alkaline Battery (A002), Protective Boot w/ magnets (A614).
The TPI 665 is a Dual Input Differential Manometer with ±100 psi range and 0.01 psi resolution. Data logging with optional A665 software and docking station. Seven units of pressure: inH2O, mbar, KPa, PSI, mH2O, mmHg, inHg, Auto power off, Record/Min/Max, Backlight display, Zero function, Trim mode to stabilise readings. ¼” NPT fittings.
Logging Data
This mode sets up the interval or time between each logged data point. The interval can be set from 1 second to 24 hours. The TPI 665 can log a total of 5000 data points. NOTE: The optional A665 docking station and software is required to view logged information.
FEATURES
Differential input with +/-100psi range.
Seven units of measure (bar, kPa, PSI, mmHg, inHg, mH2O, inH2O).
Data Logging - Record test data with time stamp and display with optional software.
PC Communication: IR to USB Interface and PC Application software (optional p/n A665).
Print test results directly to the optional A740 IR printer.
Display - 5 digit dual line with backlight.
Data Storage - Save up to 16 readings.
Min / Max - Records Min/Max readings.
Auto Off - Conserve batteries.
Measurement Range | ±101.5 psi (±7bar). |
Display | 5 Digit Dual line LCD with backlight. |
Input Type | Differential (1/4” NPT). |
Resolution (max) | 0.01 psi (0.001 bar) (0.1 inH2O). |
Accuracy | 0.2% Full scale. |
Functions | One Touch Zero, Data Hold, Back-light, Unit Selection, Smoothing, Auto power Off, Min / Max Record, Resolution Selection. |
Units of Measure | 7Units (bar, kPa, PSI, mmHg, inHg, mmH2O, inH2O). |
Data logging | 5000 Data, 1sec ~ 24 hour. |
Data Storage | 16 data points. |
Communication | IR to USB Interface and PC Application software (optional p/n A665). |
Battery Type | AA(1.5V)*2 alkaline (MN1500, LR6). |
Battery Life | 600 hours (Typical). |
Auto Power Off | 20 minutes (can be disabled). |
No customer reviews for the moment.