- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
Fluke 719-30G Electric Pressure Calibrator
Includes: Protective Yellow Holster, Instructions Sheet (15 languages), Test Leads and Alligator Clip, Two 9V alkaline batteries, Hose Kit, NIST traceable calibration report and data.
Fluke 719-30G Electric Pressure Calibrator, 30 psi, 2 bar.
With the innovative, built-in electric pump, the Fluke 719 Electric Pressure Calibrator provides pressure calibration at your fingertips.
Use the programmable pump limit settings to eliminate over-pressurisation, and to set ‘pump to’ pressure values. The unique cleanout port design reduces pump failures, and makes the Fluke 719 easy to clean without repair. The compact, lightweight form, combined with one-handed calibration, makes this pressure calibrator the ideal tool for busy process professionals.
The Calibrator makes 5-digit pressure measurements in the following units: Psi, inH2O at 4°C, inH2O at 20°C, kPa, cmH2O at 4°C, cmH2O at 20°C, bar, mbar, kg/cm2, inHg, mmHg.
The Fluke 719 displays pressure and current measurements simultaneously. The upper display shows the applied pressure or vacuum (shown as a negative value). The lower section of the display shows the current (up to 24 mA) applied to the current (mA) inputs, or the mA output value.
FEATURES
Electric pump for one handed pressure calibration.
Best in class pressure measurement uncertainty of 0.025%.
Measure and source mA with best in class 0.015% accuracy.
Precision pressure adjust vernier for accurate pressure calibration.
Variable release rate bleed valve for easy pressure adjustment.
Ideal performance for high accuracy transmitter calibration.
Programmable pressure calibrator pump limit.
Extended pressure measurement range with any of 29 700Pxx Pressure Modules.
Source mA with simultaneous pressure measurement to test valves and I/Ps.
Simulate mA signals to troubleshoot 4-20 mA loops.
Test pressure switches with built in switch test function.
Power transmitters during test using 24 V loop supply with simultaneous mA measurement.
Pressure Calibrator includes proven cleanout ports reduce pump failures.
General Specifications | ||
Storage temperature | -30°C to 60°C. | |
Operating temperature | -10°C to 55°C. | |
Relative humidity | 95% (-10°C to 30°C). | |
75% (30°C to 40°C). | ||
45% (40°C to 50°C). | ||
35% (50°C to 55°C). | ||
Shock | 1 meter drop test. | |
Safety | CSA C22.2 No. 1010.1:1992. | |
EMC | EN50082-1:1992 and EN55022:1994 Class B. | |
Size (H x W x L) | 60 x 87 x 210 mm. | |
(2.19 x 3.41 x 8.28 in). | ||
Weight | 912 g (2 lb). | |
Battery | Two 9 V alkaline batteries. | |
Battery life | 12 hours with 12 mA into 500 Ω, 200 pump cycles to 100 psi, 500 pump cycles to 30 psi. | |
Fluke 719 warranty | Three-years, one year for pump. | |
Specifications | ||
Fluke-719 30G | Range | -12 to 36 psi, -850 mbar to 2.4 bar. |
Resolution | 0.001 psi, 0.1 mbar. | |
Accuracy | 0.025% range for 6 months 0.035% of range for one (1) year. |
|
Fluke-719 100G | Range | -12 to 120 psi, -850 mbar to 8 bar. |
Resolution | 0.01 psi, 1 mbar. | |
Accuracy | 0.025% range for 6 months 0.035% of range for one (1) year. |
|
mA | Range | 0 to 24 mA. |
Resolution | 0.001 mA. | |
Accuracy | .015% Rdg + 2 counts. | |
Max load, 1000 Ω, 750 Ω ohm in HART mode. | ||
Loop supply | Range | 24 V DC. |
Resolution | N/A. | |
Accuracy | ±10%. | |
Pump Specifications | ||
Maximum pressure generation | 120 psi (8 BAR) (100 psi model). | |
Pump to | -30 psi in less than 15 seconds1 | |
-100 psi in less than 45 seconds1 | ||
1. With 1 meter of 1/8 inch test hose. | ||
Pump cycles between rebuilds | 5000. | |
Total pump life | 20,000 calibration cycle. |
No customer reviews for the moment.
COMPONENT |
DESCRIPTION | PART No. | ||
700HTH-1 | Hydraulic Test Hose. | 2811795 | ||
700ILF | In-Line Filter. | 1566730 | ||
700PMP | Pressure Pump (10 bar). | 617131 |