loading spinner
Fluke 729 300G Automatic Pressure Calibrator
  • Fluke 729 300G Automatic Pressure Calibrator
  • Fluke 729 300G Automatic Pressure Calibrator
  • Fluke 729 300G Automatic Pressure Calibrator

Fluke 729 300G Automatic Pressure Calibrator

fluke Approved distributor

lead time   Lead Time Approx. 7-14 Days.


Includes: Fluke 729 Calibrator without wireless communication, soft case, stackable test leads, AC280 hook clips, alligator clips, test probes, magnetic hanger kit, carrying strap, quick start guide, safety sheet, power supply, rechargeable battery (installed), hose kit with push fit adapters, hanging clip, USB cable, 1/8 in FNPT to 1/4 in FNPT adapter, 1/8 in FNPT to F-M20 adapter, 1/8 in FNPT to 1/4 in FBSP adapter, PTFE tape, (2) sintered filters, water trap drain and valve tuning hoses.

4856976
£9199.00
£11038.8 Inc. VAT
Fluke 729-300G - 300 psi, 20 bar range, no wireless communication.

The Fluke 729 Automatic Pressure Calibrator has been designed specifically with process technicians in mind to simplify the pressure calibration process and provide faster, more accurate test results. Technicians know that calibrating pressure can be a time consuming task, but the 729 makes it easier than ever with an internal electric pump that provides automatic pressure generation and regulation in an in an easy-to-use, rugged, portable package.

The ideal portable pressure calibrator, the 729 allows you to simply type in a target pressure, and the calibrator will automatically pump to the desired set-point. Then, internal fine adjustment control automatically stabilises the pressure at the requested value.

The Fluke 729 can also automatically test multiple pressure test points and automatically document the results. Calibration is as easy as typing in the starting and ending pressure and the number of test points and tolerance level. The 729 does the rest by documenting the applied pressure, measured mA and % error for each test point. The bright color graphical display will flag out of tolerance test results in red.

HART communication enables mA output trim, trim to applied values and pressure zero trimming of HART pressure transmitters. You can also do light configuration tasks such as changing a transmitter tag, measurement units and ranging. Other supported HART commands include setting fixed mA outputs for trouble-shooting, read device configuration and variables and read device diagnostics.

Upload and manage documented calibration results with DPCTrack2™ Calibration Management Software, making it easy to manage your instrumentation, create scheduled tests and reports, and manage calibration data.

FEATURES

Automatic pressure generation and regulation to 300 psi.
Easily document the process using onboard test templates.
Automatic internal fine-pressure adjustment.
Measure, source and simulate 4 to 20 mA signals.
24V loop power for powering transmitters for tests.
Bright dual/triple-channel color graphical display.
HART communication for testing HART smart transmitters.
Built in leak testing for verifying test setups for leakage.
Rugged portable design with standard 3-year warranty.
Fluke Connect® App compatibility (dependent on model).
Measure temperature with optional 720RTD probe.
Measure Vdc for testing 1 V dc to 5 V dc output transmitters.
Compatible with 700 Series and 750 Series pressure modules.
Upload documented test results to DPCTrack2™ Calibration Management Software.
Innovative semi-rigid carrying case designed for field work.

Fluke 729 Product Video

Fluke 729-300G Specifications
Pressure
BAR range -0.8273 to 20.6843 bar.
PSI range -12.000 to 300.000 psi.
kPa range -82.73 to 2068.43 kPa.
Comment No wireless communication.
Electrical Specification
All specifications are valid to 110% of range, except 24 mA source and simulate which are valid to 100% of range.
Ranges
mA measure, source and simulate 0 mA to 24 mA.
Volts dc measure 0 V dc to 30 V dc.
Resolution
mA dc source, simulate and measure 1 μA.
Voltage dc measurement 1 mV.
Accuracy 0.01% ±2 LSD all ranges (at 23 °C ±5 °C).
Stability 20 ppm of full scale /°C from -10 °C to +18 °C and 28 °C to +50 °C.
mA simulate external voltage requirement 12 V dc to 30 V dc.
Loop compliance voltage 24 V dc @ 20 mA.
mA simulate external voltage requirement 12 V dc to 30 V dc.
Temperature measurement only/100 Ω Pt(385) RTD -50 °C to 150 °C (-58 °F to 302 °F).
Temperature resolution 0.01 °C (0.01 °F).
Temperature accuracy ±0.1 °C (0.2 °F) ±0.25 °C (±0.45 °F) combined uncertainty when using 720 RTD probe (optional accessory).
Drive capability 1200 Ω without HART resistor, 950 Ω with internal HART resistor.
General
Operating temperature -10 °C to 50 °C for measurement, 0 °C to 50 °C for pressure control.
Operating temperature with battery -10 °C to 40 °C.
Battery will only charge to 40 °C.
Storage temperature -40 °C to 60 °C.
Dimensions (H x W x L) 7 x 27.94 x 17.27 cm (2.75 x 11 x 6.8 in).
Weight 6.5 lbs., 2.95 kg.
Enclosure rating IP54.
Fluke 729-150g warranty Two-years (standard), extended warranties are available.

Download

Fluke 729 Datasheet

PDF Specifications of the Fluke 729 Automatic Pressure Calibrator.

Download (1.15M)

No customer reviews for the moment.

Write your review

Fluke 729 300G Automatic Pressure Calibrator
COMPONENT
DESCRIPTION PART No.
BP729 Rechargeable Li-ion Battery. 4817068
Charger Power Charger, AC/DC. 4878453
AC280 Suregrip Hook Clip, Black. 4253130
AC280 Suregrip Hook Clip, Red. 4253148
TP220 Test Probes, Red. 4253345
TP220 Test Probes, Black. 4253338
Hose Nylon Hose 3.3 Ft. 4366602
Fitting 1/8" NPT-Female X 1/4" NPT-Female. 4366616
Fitting 1/8" Tube x 1/8" NPT-Male. 4551693
Fitting 1/8" NPT-Female x M20-Female. 4366633
Fitting 1/4" BSP-Female x 1/8" NPT-Female. 4366640
Case Softcase Accessory. 4821227
Trap Liquid Contaminant Trap (optional). 4380747

16 other products in the same category: