- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
Ametek ETC-400A Easy Temperature Calibrator
Includes: Jofra ETC-400 A series dry-block, 230 VAC power, power cord, metric drilled multi-hole block, standard NPL/NIST traceable certificate and carrying case.
The minimal mass offers an extremely fast heating and cooling time. The different layouts also make it possible to use an external JOFRA STS reference probe during the calibration. Choose the combination of holes that best suits your needs from our various design combinations.
The ETC series has a very easy and straight forward procedure for recalibration/adjustment. There is no need for a screwdriver or PC software. The only thing you need is a reliable reference thermometer. Place the probe in the calibrator and follow the instructions on the display.
NOTE: All ETC400 calibrators are with fixed inserts. They can NOT be changed.
ETC SERIES FEATURES
Temperature range from -10 to 400°C (14 to 752°F) with 3 models.
Accuracy to ± 0.5°C and Stability to ± 0.05°C.
Time saving fast cooling and heating times.
Easy-to-read multi-information display.
Customizable stability checkmark with countdown timer.
Set exact temperature desired to a resolution of 0.1°C.
ETC 400R, small lightweight IR calibrator.
Lightweight and easy to carry.
Temperature range @ ambient temp. 23°C / 73°F | |
---|---|
Range | 28 to 400°C / 82 to 752°F@ 23°C. |
Resolution |
|
Selectable | 1º or 0.1º. |
Heating Time | |
28 to 200°C / 82 to 392°F 200 to 400°C / 392 to 752°F |
2 minutes 3 minutes. |
Cooling Time | |
400 to 200°C / 752 to 392°F 200 to 50°C / 392 to 122°F |
6 minutes 15 minutes. |
Stability | ±0.15°C / +0.27°F. |
Measured after the stability indicator has been on for 10 minutes. Measuring time is 30 minutes. | |
Time to stability | 3 minutes (approximate). |
Accuracy | +0.5°C / +0.9°F. |
General Specifications | |
Immersion depth | 105 mm / 4.1 in. |
Voltage AC | 115V (90-132) and 230V (180-264). |
Frequency | 45 - 65 Hz. |
Power consumption (max.) | 350 W. |
Serial data interface | RS232. |
Operating temperature | 0 to 40°C / 32 to 104°F. |
Storage temperature | -20 to 50oC / -4 to 122°F. |
Humidity | 0 to 90% RH. |
Protection class | IP10. |
DNV Marine Approval, Certificate no | A-10384. |
Dimensions (L x W x H) | 172 × 72 × 182 mm / 6.8 × 2.8 × 7.2 in. |
Weight | 1.6 kg / 3.5 lb. |
Ametek ETC-400A warranty | Two-years. |
No customer reviews for the moment.
COMPONENT | DESCRIPTION | PART No. | ||
Brushes - 4 mm | Cleaning Brushes - 4 mm - Package of 3 pcs. | 122832 | ||
Brushes - 6 mm | Cleaning Brushes - 6 mm - Package of 3 pcs. | 60F174 | ||
Brushes - 8 mm | Cleaning Brushes - 8 mm - Package of 3 pcs. | 122822 | ||
Converter | Edgeport Converter with 4 pcs of RS232 ports. | 125002 | ||
Case | Carrying Case for ETC Series. | 124094 |