- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
Seaward Sentinel 500 Electrical Safety Tester
Includes: Clare Sentinel 500 User Manual, test lead set (hv cables x 2), interlock key, remote terminal plug, power lead.
The Sentinel 500 electrical safety tester enables manufacturers to check their products in-house to ensure compliance with the relevant international technical conformance standards such as EN 60598, UL 1598, IEC and CSA C22.2 before being sent to third party test/approval houses. This can reduce product time-to-market, saving time and costs in the process.
The fast, safe and accurate testing of high voltages up to 5kV AC and 6kV DC is provided across the range while accurate testing up to 500 VA AC power output - allowing for 100mA tripping currents, a requirement in many standards - is available on the Sentinel 500 and 501 units
A remote start terminal and signal I/O port allows for remote starting, safety interlocking and PLC control while an enhanced safety feature will automatically cut off the output voltage within 150µs where normal output voltages are detected or when the upper or lower current trip limits are reached while the device is under test.
The Clare Sentinel series is ideal for R&D facilities and compliance laboratories, while design, quality assurance, and manufacturing and production engineers will also benefit from the easy-to-use safety test features.
SENTINEL SERIES KEY FEATURES500 VA capability (Sentinel 500 series).
Earth bond / insulation resistance (Sentinel 501).
Sweep function for test analysis.
Flexible / high resolution test parameters .
Manual / automatic test modes.
Remote and PLC control.
Fast cut off and automatic discharge.
Features | 200 | 500 | 501 |
---|---|---|---|
Earth/ground Bond | |||
AC dielectric strength - 5 kV | |||
DC dielectric strength - 6 kV | |||
DC insulation resistance | |||
200 VA (40 mA) | |||
500 VA (100 mA) | |||
ARC detection |
AC Dielectric Strength | ||
Programmable Voltage Range | 0.100 kV - 5.000 kV (2V/Step Resolution). | |
Frequency Independent of Supply | 50 or 60 Hz Selectable. | |
Output Voltage Accuracy | ±1% + 5 Volts (No Load). | |
Voltage Display Resolution | 0.001 kV. | |
Current Measurement | 0.001 mA - 100.0 mA. | |
Range and Accuracy | ±1.5% + 30 Counts when trip limit <1.11mA ±1.5% ± 3 Counts when trip limit >1.11mA. |
|
Current Display Resolution | 0.001 mA / 0.001 mA / 0.1 mA. | |
Selectable Range of Pass/Fail Levels | 0.001 mA - 100.00 mA. | |
Maximum Rated Current | 100 mA (0.5 kV ≤ V ≤ 5 kV) 10 mA (0.1 kV ≤ V ≤ 0.5 kV). |
|
Arc Detection | Yes. | |
Ramp time | 0.1 - 999.9s. | |
Test Time | Off / 0.5s - 999.9s. | |
DC Dielectric Strength | ||
Programmable Voltage Range | 0.100 kV - 6.000 kV (2V/Step Resolution). |
|
Frequency Independent 50 or 60 Hz Selectable of Supply | 50 or 60 Hz Selectable. | |
Output Voltage Accuracy | ±1% + 5 Volts (No Load). | |
Voltage Display Resolution | 0.001 kV. | |
Current Measurement | 0.001 mA - 20.0 mA. | |
Range and Accuracy | ±1.5% + 30 Counts when trip limit <1.11 mA ±1.5% + 3 Counts when trip limit >1.11 mA. |
|
Current Display Resolution | 0.001 mA / 0.001 mA / 0.1 mA. | |
Selectable Range of Pass/Fail Levels | 0.001 mA - 20.00 mA. | |
Maximum Rated Current | 20 mA (0.5 kV ≤ V ≤ 5 kV) 2 mA (0.1 kV ≤ V ≤ 0.5 kV). |
|
Arc Detection | Yes. | |
Ramp time | 0.1 - 999.9s. | |
Test Time | Off / 0.5s - 999.9s. | |
General Specifications | ||
Dimensions | 330W x 148H x 587D mm. | |
Weight | Approx 27 kg. | |
Sentinel 500 warranty | 1 year (subject to product registration with Seaward). |
No customer reviews for the moment.