Chauvin C.A 6255 Microhmmeter
Includes: 1 set of cables 3 m long terminated by Kelvin clamps, mains power lead, 9 user’s manuals (1 per language), 9 simplified user’s manuals (1 per language), MOT (Micro-Ohmmeter Transfer) data transfer software on CD-ROM, 1 x RS232 communication cable.
An easy-to-use, high-performance instrument, the Chauvin C.A 6255 10A micro-ohmmeter tests the quality of connections with precision (resolution of 0.1 µΩ) through the use of low resistance measurements. Housed in a rugged construction-site type case with cover, the C.A 6250 is a self-contained instrument powered by a rechargeable battery with built-in charger.
It provides 7 measurement ranges, from 5 mΩ to 2,500 Ω, that can be accessed and selected directly on the front-panel rotary switch. It uses the 4-wire measurement method, with automatic compensation of spurious voltages.
This device is compliant with safety standard IEC 61010-2-030 and the leads are compliant with IEC 61010-031, for voltages up to 50 V with respect to earth in category III.
3 Measurement Modes
NON-INDUCTIVE MODE - for measurements on resistors with a time constant of a few ms (contacts, metallisation, etc.).
INDUCTIVE MODE - for measurements on inductive components (transformers, motors, etc.).
NON-INDUCTIVE MODE - with AUTOMATIC trigger: for measurements on resistors without a time constant: for repetitive measurements in production, helping to gain considerable time.
Temperature Compensation
The value of a resistance varies according to the temperature. For reliable supervision allowing the comparison of measurements which are comparable because they were made in the same conditions, it is a good idea always to express a measurement in terms of a given reference temperature. You can perform this calculation by simply pressing a key.
FEATURES
Automatic detection of the presence of an external AC or DC voltage on the terminals
3 different measurement modes depending on the nature of the resistance to be measured,
Protection of the operator when a resistance having a large inductive component (motor, transformer, etc.) is measured.
Measurement up to 2,500 Ω, resolution 0.1 μΩ.
Programming of alarm thresholds.
Measurement temperature using a Pt100 jack on the front panel,
Function for automatic calculation of the resistance at a reference temperature
Extended memory making it possible to store approximately 1,500 measurements,
Indication of the level of memory use.
Indication of battery charge condition.
Automatic switching of the backlighting to standby to save battery power,
RS 232 interface to print the results on a serial printer or export them to a PC.
4-wire measurement with compensation of spurious voltages. | |||||
RANGE | RESOLUTION | ACCURACY OVER 1 YEAR |
MEASURING CURRENT |
VOLTAGE DROP |
|
5.000 mΩ
|
0.1 µΩ | 0.05% + 1.0 µΩ | 10 A | 50 mV | |
25.000 mΩ | 1 µΩ | 0.05% + 3 µΩ | 10 A | 250 mV | |
250.00 mΩ | 10 µΩ | 0.05% + 30 µΩ | 10 A | 2500 mV | |
2500.0 mΩ | 0.1 mΩ | 0.05% + 0.3 mΩ | 1 A | 2500 mV | |
25.000 Ω | 1 mΩ | 0.05% + 3 mΩ | 100 mA | 2500 mV | |
250.00 Ω | 10 mΩ | 0.05% + 30 mΩ | 10 mA | 2500 mV | |
2500.0 Ω | 100 mΩ | 0.05% + 300 mΩ | 1 mA | 2500 mV | |
Measurement modes | Inductive, non-inductive, non-inductive with automatic trigger. | ||||
Temperature compensation | By temperature sensor or manual. | ||||
Data storage | 1,500 measurements. | ||||
Operating temperature | 0°C to +50°C. | ||||
Electrical safety | EN 61010-1. | ||||
Max. voltage relative to earth: | 50 V. | ||||
Mechanical protections |
IP53 = housing open.
IP64 = housing closed. |
||||
Dimensions | 270 x 250 x 180mm. | ||||
Weight approx. | 4 kg. | ||||
Chauvin C.A6255 warranty | Two years. |
No customer reviews for the moment.
COMPONENT |
DESCRIPTION | PART No. | ||
10A Clamps | 10 A Kelvin clamps 3 m (set of 2). Spade lug. | P01101794 | ||
10A Probes | Double 10 A “pistol” test probe (set of 2) L = 3.15 m, Ø 2 mm test probes. | P01103065 | ||
10A Probes | Double 10 A pivoting test probe (set of 2) L = 3.15 m, Ø 3 mm test probes. | P01103063 | ||
1A Clamps | 1 A Mini Kelvin clamps (set of 2), Spade lug. | P01101783 | ||
1A Probes | Double 1 A test probe (set of 2) L = 2.85 m, Spade lug and 4 mm banana. | P01102056 | ||
PT100 | Temperature probe. | P01102013 |