- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
Fluke 64 MAX Infrared Thermometer
Includes: IR Thermometer and Printed instruction sheet.
The Fluke 64 MAX IR thermometer has the precision you need and won't break your budget. Designed & tested to survive a 3m drop, this tool works in harsh environments, even when you’re not there.
The Fluke 64 MAX displays minimum, maximum, or average temperature, or the difference between two measurements, and features Hi and Lo alarms for rapid display of measurements outside set limits. It can survive a three-meter drop and continue to operate, making it ideal for industrial and field environments.
AUTO-CAPTURE
The Fluke 64 MAX IR Thermometer’s Auto Capture feature allows a user to capture multiple measurements unattended, when they are in another area of the plant or in the middle of the night after they have gone off shift. Just set the time to start the measurements and the interval you want to capture them. Then place the 64 MAX on a standard tripod with the tripod mount (offered separately). Set it up and walk away.
FEATURES
Precise laser technology for more accurate and repeatable measurements.
Temperature accuracy of up to ± 1℃ or ± 1% of reading whichever is greater with 20:1 distance to spot ratio.
Flashlight and large, easy-to-read backlit LCD display for easy viewing even in dark environment.
IP54 rated for extra protection against airborne contaminants.
Set time and desired interval between measurements and Auto Capture will capture spot temperatures unattended.
99 data point logging.
30 hour battery life so your tool can perform as long as you need it to.
Displays the minimum, maximum or average temperature, or the difference between two measurements.
Hi and Lo alarms for rapid display of measurements outside set limits.
Small and lightweight design fits easily into your tool box.
Powered by one (1) AA alkaline battery.
Temperature Range | -30 °C to 600 °C (-22 °F to 1112 °F). |
Accuracy (Calibration geometry with ambient temperature 23 °C +/- 2 °C | ± 1.0 °C or ± 1.0 % of reading, whichever is greater ± 2.0 °C at -10 °C to 0 °C ± 3.0 °C at -30 °C to -10 °C. |
Optical Resolution | 20 : 1 (calculated at 90 % energy). |
Response Time (95 %) | < 500 ms (95% of reading). |
Spectral Response | 8 to 14 microns. |
Emissivity | 0.10 to 1.00. |
Display Resolution | 0.1 °C (0.2 °F). |
Repeatability (% of reading) | ± 0.5% of reading or ± 0.5 °C (whichever is greater). |
Auto Capture | Set time and interval to capture up to 99 measurements. |
Internal memory | 99 data point logging capability. |
Flashlight | Yes. |
Power | 1 AA battery. |
Battery life | 30 hours with laser and backlight on. |
Weight | 255 g (8.99 oz). |
Dimensions | (175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches. |
Operating Temperature | 0 °C to 50 °C (32 °F to 122 °F). |
Storage Temperature | -20 °C to 60 °C (-4 °F to 140 °F) (without battery). |
Operating Humidity | Non- condensing @ ≤ 10 °C (50 °F) ≤ 90 % RH @ 10 °C (50 °F) to 30 °C (86 °F) ≤ 75 % RH @ 30 °C (86 °F) to 40 °C (104 °F) ≤ 45 % RH @ 40 °C 104 °F to 50 °C (122 °F). |
Operating Altitude | 2,000 meters above mean sea level. |
Optical Resolution | 12,000 meters above mean sea level. |
Drop test | 3 meters. |
Implement Standard | Q/ASF01. |
Compliance | IEC 61010-1: Pollution Class 2. |
Laser Safety | IEC 60825-1 Class 2, 650 nm, < 1 mW |
Ingress Protection Rating | IP 54 per IEC 60529. |
Fluke 64 Max Warranty | 3 years. |