Megger PAT320 PAT Tester
Includes: Carry case, Quick-start guide, Black test lead set with probe and clip, IEC lead 0.5 m (Extension lead adaptor), Warranty card, Owners CD manual, PP3 9 V NiMH rechargeable battery.
The Megger PAT320 is the PAT tester if you need high productivity with a tough and easy to use instrument. If you are moving into PAT testing for the first time or you don’t want the complexity of database management, PAT320 is ideal for you.
Simple push-button operation make the Megger PAT320 very fast and easy to use. All regulatory test requirements are supported, including Class I and Class II tests. IEC power leads and extension leads are easily connected. An automatic mode is available for Class I and Class II testing.
FEATURES
Simple manual or automatic testing.
230V and 110V operation.
Virtual continuous operation.
Bond testing at 25 A,10 A and 200 mA.
Full colour display.
Tests portable RCDs and extension leads.
Differential, touch and substitute leakage testing.
Electrical Supply Range |
||
230 V AC | 230 V AC ±10%. | |
110 V AC | 110 V AC ±10%. | |
Supply Measurement | ||
Voltage | ±2% ± 1 V | |
Frequency | ±1% ± 0.1 Hz. | |
Bond Test | ||
Open Circuit Voltage | 9 V AC ±10% ±0.1 V (Supply: 230 V 50 Hz). | |
10 A Bond Test Current | 10 A rms ±20% ±0.5 A into 0.1 Ω. | |
25 A Bond Test Current | 25 A rms ± 5% ± 0.5 A into 0.1 Ω. | |
Earth Bond Resistance Accuracy | ±5% ± 3 digits (0 to 0.49 Ω) ±5% ± 5 digits (0.5 to .99 Ω). |
|
Earth Bond Resistance Resolution | 10 mΩ (0 to 1.99 Ω). | |
Display Range | 0 to 1.99 Ω. | |
Bond Test Nulling | Up to 1.00 Ω. | |
Adjustable Test Duration | User selectable from 1 sec to 20 sec. | |
Continuity Test | ||
Continuity Test Voltage | 4.0 V DC -0% +10% (open circuit). | |
Continuity Test Current | 200 mA -0% + 10% ± 5 mA (into 2 Ω load). | |
Continuity Resistance Accuracy | ±5% ±3 digits (0 to 0.49 Ω) ±5% ±5 digits (0.5 to .99 Ω). |
|
Continuity Resistance Resolution | 10mΩ (1 to 19.99 Ω). | |
Display Range | 0 to 19.99 Ω. | |
Continuity Test Nulling | Up to 9.99 Ω. | |
Test Duration | User selectable from 1 sec to 20 sec. | |
Note: The continuity test is a DC test performed automatically in both positive and negative directions. The average of the two results is shown. | ||
Insulation Test | ||
Insulation Test Voltage | 250 V DC -0%/+25% open circuit 500 V DC -0%/+25% open circuit (500 V DC across 0.5 MΩ). |
|
Insulation Resistance Accuracy (230 V) | ±2% ±5 digits (0 to 19.99 MΩ) ±5% ± 10 digits (20 to 99.99 MΩ). |
|
Insulation Resistance Accuracy (110 V) | ±5 ±5 digits (0 to 10 MΩ). | |
Insulation Resistance Resolution | 0.01 MΩ (0.10 to 99.99 MΩ). | |
Display Range | 0 to 99.99M Ω. | |
Test Duration | User selectable from 1 sec to 1 minutes. | |
Substitute Leakage Test | ||
Test Voltage | 40 V AC ±10%. | |
Test Frequency | Nominal mains 50/60 Hz. | |
Leakage Current Accuracy | ±5% ± 5 digits. | |
Leakage Current Resolution | 0.01 mA. | |
Display Range | 0 to 19.99 mA. | |
Test Duration | User selectable from 1 sec to 1 minute Reading corrected to 230 V + 10% or 110 V +10% AC. |
|
Differential Leakage Current | ||
Test Voltage | Nominal mains 110 V/230 V AC. | |
Test Frequency | Nominal mains 50/60 Hz. | |
Differential Leakage Current Accuracy | ±5% ± 5 digits. | |
Differential Leakage Current Resolution | 0.01 mA. | |
Display Range | 0 to 19.99 mA Reading corrected to 230 V +10 % or 110 V + 10% AC. |
|
Touch Current Test | ||
Test Voltage | Nominal mains 115 V/230 V AC. | |
Test Frequency | Nominal mains 50/60 Hz. | |
Touch Current Accuracy | ±5% ±5 digits. | |
Touch Current Resolution | 0.01 mA. | |
Display Range | 0 to 10 mA. | |
Test Duration | User selectable from 1 sec to 5 sec Reading corrected to 230 V + 10% or 110 V + 10% AC. |
|
Operational Test | ||
Test Voltage | Nominal mains 110 V/230 V AC. | |
Accuracy | ±5% ±10 digits (0 to 99 VA) ±5% ±50 digits (100 to 999 VA) ±5% ±100 digits (1000 to 3700 VA). |
|
Resolution | 1 VA (0 to 3700 VA). | |
Display Range | 0 to 3.99 KVA Reading corrected to 230 V or 110 V AC Results show load VA. |
|
Extension Lead Test | ||
Polartity Test Voltage | 12 V | |
Polarity | Leak OK Live neutral S/C Live neutral reversed Live/neutral O/C. |
|
Portable RCD Test | ||
RCD Test Voltage | Nominal mains 110/230 V. | |
RCD Test Frequency | 50 Hz. | |
Test Current Accuracy | -8% to -2% (1/2 x I) ±2% to +8^ (1 x 1,5 x I). |
|
Trip Time Accuracy | ±1% ±5 digits. | |
Trip Time Resolution | 0.1ms. | |
Display Range | 0 to 1999 ms (1/2 x I) 0 to 300 ms (1 x I) 0 to 40 ms (5 x I). |
|
Fuse Test | ||
Test Voltage | 3.3 V. | |
Warning | Audible Beep if fuse is ok. | |
Dimensions (with case) | 190 (H) x 400 (D) x 290mm (W). | |
Weight (instrument only) | 4.5 kg. | |
Megger PAT320 warranty | 1 year. |
No customer reviews for the moment.
COMPONENT |
DESCRIPTION | PART No. | ||
110 V Adapt | 110 V mains lead adaptor. | 1000-766 | ||
Plug Adapt | Plug adaptor IEC C6 - C13. | 2000-551 | ||
Ext lead | 110 V Extension lead adaptor. | 6220-639 | ||
Adapt | 415 V adaptor lead (4-pin) to BS1363 (16 A). | 1000-767 | ||
Adapt | 415 V adaptor lead (5 pin) to BS1363 (16 A). | 1000-770 | ||
Protect | Screen protector (2). | 1002-572 | ||
Pass Labels | Roll of 1000 PASS test labels. | 1000-971 | ||
Fail Labels | Roll of 1000 FAIL test labels. | 1001-227 |