The Fluke 707 Loop Calibrator is a compact sourcing and measuring tool. The Calibrator tests current loops of 0-20 mA or 4-20 mA and measures dc voltage to 28 V.

The 705 Calibrator is a IEC 61010, CAT I 30 V, Pollution Degree 2 instrument. A CAT I instrument is designed to protect against transients from high-voltage, low-energy sources, like electronic circuits or a copy machine, for example.

The Fluke 707 calibrator outputs current for calibrating and testing 0 to 20 mA and 4 to 20 mA current loops and instruments. In SOURCE mode, the Calibrator supplies the current. In SIMULATE mode, the Calibrator simulates a 2-wire transmitter in an externally-powered current loop.

FEATURES:

Large display and simple, quick-click interface for easy one-handed operation.
Simultaneous mA and % readout for quick, easy interpretation of readings.
mA accuracy of 0.015%, superior to other loop calibrators in this price range.
1 µA resolution for mA source, simulate and measure.
Push button 25% steps for fast, easy linearity checks.
0-100% “span check” for fast confirmation of zero and span.
Selectable slow ramp, fast ramp, and step ramp to provide smooth outputs for valve slewing and loop functional tests.
24 volt internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM.
Measures V dc to 28 V.
Single 9V battery that is easily changed without breaking calibration seal.
0 - 20 mA or 4 - 20 mA default start up modes.
Innovative output adjustment dial with 1 µA and 100 resolution.
HART ™ mode connects 250 Ω resistor in series with 24 V loop for compatibility with HART communicators.

Fluke 707 Specifications
Current measurement Range 0 to 24 mA.
Resolution 0.001 mA.
Accuracy 0.015% Rdg + 2 LSD.
Current sourcing Range 0 to 20 mA or 4 to 20 mA.
Accuracy 0.015% Rdg + 2 LSD.
Drive capability 1200 Ω 950 Ω at 20 mA in HART mode.
Current simulation Range 0 to 20 mA or 4 to 20 mA.
Accuracy 0.015% Rdg + 2 LSD.
Ext. loop voltage requirement 12-30 V.
Loop power While measuring mA, 24 V.
Voltage measurement Range 0 to 28 V DC.
Resolution 1 mV.
Accuracy 0.015% Rdg + 2 LSD.
Environmental Specifications
Operating temperature -10°C to 55°C.
Storage temperature -30°C to 60°C.
Humidity (without condensation) 0 to 90% (0°C to 35°C); 0 to 70% (35°C to 55°C).
Temperature coefficient -10°C to 18°C, 28°C to 55°C, ±.005% of range per °C.
Safety Specifications
Safety CSA C22.2 No. 1010.1:1992.
EMC EN50082-1:1992 and EN55022:1994 Class B.
Maximum voltage 30 V.
Mechanical and General Specifications
Size 152 x 70 x 41 mm.
Weight 224 g (340 g including holster).
Batteries One 9 V battery Alkaline.
Warranty Three years.
Shock and vibration Per MIL-T-28800 for a Class 2 Instrument.
Display current and % of span Yes.
Auto step, auto ramp Yes.
Span check Yes.

Download

Fluke 705-707-715 Datasheet

PDF Datasheet of the Fluke 705, Fluke 707 & Fluke 715 Loop Calibrators.

Download (1.28M)

No customer reviews for the moment.

Write your review

Fluke 707 Loop Calibrator
COMPONENT PARTS DESCRIPTION PART No.
AC220 SureGrip™ Alligator Clip Set. 1610103
TL220-1 SureGrip™ Industrial Test Lead Set includes:
AC220 SureGrip™ Alligator Clip Set
TP220 SureGrip™ Industrial Test Probe Set
TL224 SureGrip™ Silicone Test Lead Set (straight to right)
3971178

16 other products in the same category:

arrow_upward