- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
Fluke 718Ex 300G IS Pressure Calibrator
Includes: Fluke 718Ex 300G, Protective red holster, TL75 test leads, AC72 test clips, Media filter, CD users manuals (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, Japanese), 718Ex CCD content control drawing.
The 718 Ex Intrinsically Safe Pressure Calibrator is a powerful intrinsically safe, self contained pressure calibrator for usage in Explosion endangered areas. The Fluke 718Ex makes 5-digit pressure readings in the following units: psi, inH²O at 4 °C, inH²O at 20 °C, kPa, cmH²O at 4 °C, cmH²O at 20 °C, bar, mbar, kg/cm², inHg, and mmHg.
FEATURES
ATEX II 1G Ex ia IIC T4 Compliant.
Built-in pressure / vacuum hand pump, with fine adjust vernier and bleed valve.
30, 100, and 300 psi ranges (2, 7, and 20 bar)
Pressure measurement to 0.05% of full span, using an internal pressure sensor.
Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 700PEx Pressure Modules.
Wide range of selectable measurement units for pressure.
Current measurement with 0.02% accuracy and 0.001 mA resolution.
Pressure switch test function.
Min-Max Hold functions.
Compact size and weight.
Simple, push-button user interface.
Rugged and reliable, ideal for field use.
Pressure Measurement Accuracy | ||
Fluke 718Ex 30G | Range | -830 mbar to 2 bar (internal sensor) Over pressure: 3x FS. |
Resolution | 0.0001 bar. | |
Accuracy | 0.05% FS. | |
Functions | Zero, Min, Max, Hold, Damp. | |
Fluke 718Ex 100G | Range | -830 mbar to 2 bar (internal sensor) Over pressure: 2x FS. |
Resolution | 0.001 bar. | |
Accuracy | 0.05% FS. | |
Functions | Zero, Min, Max, Hold, Damp. | |
Fluke 718Ex 300G | Range | -830 mbar to 20 bar (internal sensor) Over pressure: 375 psi, 25 bar. |
Resolution | 0.0001 bar. | |
Accuracy | 0.05% FS. | |
Functions | Zero, Min, Max, Hold, Damp. | |
Fluke 718Ex series | Range | Via pressure module connector 29 pressure modules, 2.5 mbar to 700 bar Over pressure per pressure module specs. |
Resolution | Per pressure module specs. | |
Accuracy | To 0.025% of full span, per pressure module specs. | |
Functions | Zero, Min, Max, Hold, Damp. | |
Source pressure, built in pump range | -820 mbar to full scale. | |
For use with non-corrosive gasses. Supported Pressure Units psi, inH2O (4°C), inH2O (20°C), kPa, cmH2O (4°C), cmH2O (20°C), bar, mbar, kg/cm², mmHg, inHg Media compatibility per pressure module specs. |
||
Technical Data | ||
Measure accuracy | Range | 0-24 mA. |
Resolution | 0.001 mA. | |
Accuracy | 0.02% rdg ±2 counts. | |
Display | LCD, 5 digit pressure and current simultaneous. | |
Environmental Specifications | ||
Operating temperature | -10°C to 55°C. | |
Storage temperature | -40°C to 60°C. | |
Humidity (without condensation) | 95% (10°C to 30°C). | |
75% (30°C to 40°C). | ||
45% (40°C to 50°C). | ||
35% (50°C to 55°C). | ||
Operating altitude | 3,000 m max. | |
Safety Specifications | ||
Safety | ATEX II 1G Ex ia IIC T4 compliant. | |
CSA Certification I.S. Class I. | ||
Div 1 Groups A-D , T4. | ||
Mechanical and General Specifications | ||
Size (L x W x H) | 216 x 94 x 66 mm (with holster). | |
Weight | Approx. 992 g (with holster). | |
Warranty | Three years, one year for pump. | |
Shock | 1 meter drop test. | |
Vibration | Random, 2 G, 5-500 Hz. | |
Battery type | One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V. | |
Battery life | 4 to 20 hours, depending on functions used. |
No customer reviews for the moment.
COMPONENT |
DESCRIPTION | PART No. | ||
700HTH-1 | The Fluke 700HTH-1 hydraulic test hose allows connection to both a Fluke pressure module and a calibration unit under test from a Fluke 700HTP-1 hydraulic test pump simultaneously. | 2811795 | ||
TL238 | SureGrip™ Industrial Test Lead Set designed to provide a reliable grip in slippery hands. With interchangeable test lead set for multiple electrical applications. | 2132831 | ||
TL71-1 | The TL71-1 Premium Test Lead Set has one pair of red and black comfort grip probes with flexible silicone-insulated, right-angle test leads.. | 3971207 |