- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
Fluke 3130-GM2 Portable Pressure Calibrator
Includes: Manual, NIST-traceable certificate, test leads, universal ac adapter/charger.
The Fluke 3130 Portable Pneumatic Pressure Calibrator is ideal for calibrating pressure transmitters, transducers, gauges and similar devices. The Fluke 3130 contains everything you need to generate, control and measure pressure, as well as read the output of the device under test (DUT).
The internal, motorised pump provides an easy, efficient alternative to time consuming, tiring hand pumps. For applications that involve filling a large volume with pressure, the Fluke Calibration 3130-G2M allows for connection to an external gas supply such as compressed plant air. The pressure can be fine-tuned using the variable volume.
The Fluke 3130 can be powered with the internal, high-capacity NiMH battery and/or the universal power supply (90 V ac to 250 V ac) that charges the battery pack. The universal power supply can power the Product at the same time it charges the battery pack. A fully discharged battery can take a maximum of 14 hours to recharge. When the calibrator is first received, plug it into the universal power supply for that period to ensure it is fully charged.
Engineering units: psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4 °C, mmH2O @ 20 °C, cmH2O @ 4 °C, cmH2O @ 20 °C, inH2O @ 4 °C, inH2O @ 20 °C, inH2O @ 60 °F, mmHg @ 0 °C, inHg @ 0 °C.
FEATURES
Measure and generate pressures from vacuum to 2 MPa (300 psi, 20 bar).
Internal pump can generate vacuum to -80 kPa (-12 psi, -0.8 bar) or pressure to 2 MPa (300 psi, 20 bar).
Supply pressure connection allowing the use of external gas supply up to 2 MPa (300 psi, 20 bar).
Includes variable volume for fine adjustment of pressures.
Pressure measurement accuracy of ± (0.025 % reading + 0.01 % FS).
Electrical measurement and 24 volt supply for close looped calibrations.
Measure or generate 4 to 20 mA.
Measure 0 to 30 V dc.
Powered by internal, high capacity, NiMH battery or universal ac mains adapter.
Compatible with Fluke 700P and 750P Pressure Modules.
APPLICATION COVERAGE
Pressure gauge calibration.
Pressure transmitter calibration.
Pressure switch test and calibration.
Environmental | ||
Operating temperature | -10 °C to +50 °C. | |
Storage temperature | -20 °C to +60 °C. | |
Power requirements | 12 V dc (Universal ac adapter/charger supplied). | |
Battery | Internal 3800 mAh advanced NiMH pack. | |
Operating Time | ||
On full charge | Approx 50 hours (Measure only or external air; no pump). |
|
Using internal pump | Approx 100 calibration cycles to 300 psi. | |
Physical | ||
Dimensions | 15.25 in L x 12 in W x 7 in D. | |
Weight | ~7 kg (15 lb). | |
EMI/RFI conformance | EN61326:2006 Annex A. | |
Connectors/ports | 1/8 in NPT (External supply port and test port). | |
Included accessories | Manual, NIST-traceable certificate, test leads, universal ac adapter/charger. |
|
Ranges | ||
Pressure (internal pump) | -80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar). | |
Pressure (external air) | 0 to 2 MPa (0 to 300 psi, 0 to 20 bar). | |
mA | 0 to 24.000 mA. | |
Volts | 0 to 30.000 V dc. | |
Engineering units | psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4 °C, mmH2O @ 20 °C, cmH2O @ 4 °C, cmH2O @ 20 °C, inH2O @ 4 °C, inH2O @ 20 °C, inH2O @ 60 °F, mmHg @ 0 °C, inHg @ 0 °C. |
|
Instrument Measurement Uncertainty | ||
Pressure | ±(0.025 % of reading + 0.01 % FS). | |
mA | ±(0.015 % of reading + 0.002 mA). | |
Volts | ±(0.015 % of reading + 0.002 V). | |
Temperature effect (all functions) | ||
No effect on accuracy on all functions from 15 °C to 35 °C. | ||
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C. |