loading spinner
Fluke 9190A-X-P Field Metrology Well With Process Electronics
  • Fluke 9190A-X-P Field Metrology Well With Process Electronics
  • Fluke 9190A-X-P Field Metrology Well With Process Electronics
  • Fluke 9190A-X-P Field Metrology Well With Process Electronics
  • Fluke 9190A-X-P Field Metrology Well With Process Electronics

Fluke 9190A-X-P Field Metrology Well With Process Electronics

fluke Approved distributor

lead time  Lead Time Approx. 14-21 Days.


Includes: 9190-INSX Insert (X=A, B, C, D, E, or F), Clamp-on Ferrites, 6-pin DIN Connector, Test Lead Kit, 2-meter (6-foot) Mains Power Cord, USB Cable, Getting Started Manual, Product CD that contains manuals and remote interface driver files, 9930 Interface-it Calibration Software and Users Guide, Report of Calibration and Calibration Label, Well Insulator Cap, Insert Removal Tool.

4178817
£20374.00
£24448.8 Inc. VAT
NB: ‘X’ in the model number to be replaced with A, B, C, D, E, and F as appropriate for the desired insert.

The Fluke 9190A ultra-cool field metrology well from Fluke Calibration provides excellent stability and uniformity and plenty of well depth for calibrating platinum resistance thermometers (PRTs), thermocouples and other industrial thermometers.

The Fluke 9190A Ultra-Cool Field Metrology Well is the most accurate and stable, cold temperature dry-block on the market. It’s ideal for applications that demand strict quality control and regulatory process compliance. These applications include on-location validation and calibration of RTDs, thermocouples, thermometers, and sensors used with process control equipment such as medical freezers, laboratory refrigerators, cold rooms, blood banks, sterilisers (autoclaves), and freeze dryers.

The Input Panel includes an input for an external reference/control thermometer that makes it possible to perform a transmitter loop calibration, comparison calibration, or a simple check of a temperature sensor. In addition, the Input Panel can read resistance and current from a probe. For probes that require power to operate, the 4-20 mA input has a loop-power function that can source up to 24 volts of power.

FEATURES


Dual heater zone: Active heater zone control compensates for differential temperatures between zones.
All controls and inputs on the front panel.
Two-handed carry; built-in handles on front and back. All controls and inputs on the front panel.
Define up to eight set-point temperatures; automatically ramp to and “soak” at each set-point.
Stability indicator: Indicates when the block temperature is stable and within limits defined by the user.
Fast time to temperature and small footprint: Uses compact, efficient free piston stirling cooler technology (FPSC).
Remote operation: USB and RS-232 communications interfaces support remote operation from a computer.
4-20 mA connectors: Power a 4-20 mA transmitter from the drywell. Supplies 24 V dc loop power (process version).
4-wire PRT/RTD input: Reads a 4-wire, 3-wire, or 2-wire PRT/RTD. Accuracy ±0.02 ºC (process version).
Reference thermometer input automatically reads Fluke Calibration PRTs terminated with a “smart” connector.
Thermocouple input: Accepts common thermocouples with mini-jack termination. Types: J, K, T, E, R, S, M, L, U, N, C.

Fluke 9190A Product Video

Fluke 9190A-X-P Specifications
Temperature range at 23 °C –95 °C to 140 °C (–139 °F to 284 °F).
Display accuracy ± 0.2 °C full range.
Accuracy with external reference3 One year.
Stability ± 0.015 °C full range.
Axial uniformity at 40 mm (1.6 in) ± 0.05 °C full range.
Radial gradient ±0.01 °C full range.
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes)
± 0.006 °C full range
(versus display with one 6.35 mm probe)
± 0.25 °C at –95 °C
± 0.10 °C at 140 °C.
Operating conditions 0 °C to 35 °C, 0 % to 90 %
RH (non-condensing) < 2000 m altitude.
Immersion (well) depth 160 mm (6.3 in).
Well diameter 30 mm (1.18 in).
Heating time1 –95 °C to 140 °C: 40 min.
Cooling time1 23 °C to –90 °C: 80 min
23 °C to –95 °C: 90 min
140 °C to 23 °C: 60 min.
Stabilisation time2 15 min.
Resolution 0.01 °.
Display LCD, °C or °F user selectable.
Computer interface RS-232, USB Serial, and 9930 interface-it temperature calibration software included.
Safety IEC 61010-1, Installation Category II, Pollution degree 2.
Electromagnetic environment IEC 61326-1: Basic.
Refrigerants R32 (Difluoromethane)
< 20 g, ASHRAE safety group A2L
R704 (Helium)
< 20 g, ASHRAE safety group A1.
Dimensions (H x W x D) 380 mm x 205 mm x 480 mm (14.9 in x 8.0 in x 18.8 in).
Weight 16 kg (35 lb).
Fluke 9190A warranty One year.
Process (-P) Specifications
Built-in reference thermometer readout accuracy (4-wire reference probe)3 ± 0.010 °C at –95 °C
± 0.013 °C at –25 °C
± 0.015 °C at 0 °C
± 0.020 °C at 50 °C
± 0.025 °C at 140 °C.
Reference resistance range 0 Ω to 400 Ω.
Reference resistance accuracy4 0 Ω to 42 Ω: ± 0.0025 Ω
42 Ω to 400 Ω: ± 60 ppm of reading.
Reference characterisations ITS-90, CVD, IEC-751, resistance.
Reference measurement capability 4 wire.
Reference probe connection 6-pin din with INFO-CON technology.
Built-in RTD thermometer readout accuracy NI-120: ± 0.015 °C at 0 °C
PT-100 (385): ± 0.02 °C at 0 °C
PT-100 (3926): ± 0.02 °C at 0 °C
PT-100 (JIS): ± 0.02 °C at 0 °C.
RTD resistance range 0 Ω to 400 Ω.
Resistance accuracy4 0 Ω to 25 Ω: ± 0.002 Ω
25 Ω to 400 Ω: ± 80 ppm of reading.
RTD characterisations PT-100 (385), (JIS), (3926), NI-120, resistance.
RTD measurement capability 2-wire, 3-wire, and 4-wire RTD with jumpers only.
RTD connection 4-terminal input.
Built-in TC thermometer readout accuracy5 Type J: ± 0.70 °C at 140 °C
Type K: ± 0.75 °C at 140 °C
Type T: ± 0.60 °C at 140 °C
Type E: ± 0.60 °C at 140 °C
Type R: ± 1.60 °C at 140 °C
Type S: ± 1.60 °C at 140 °C
Type M: ± 0.65 °C at 140 °C
Type L: ± 0.65 °C at 140 °C
Type U: ± 0.70 °C at 140 °C
Type N: ± 0.75 °C at 140 °C
Type C: ± 1.00 °C at 140 °C.
mA range Cal 4-22 mA, Spec 4-24 mA.
Loop power function 24 V dc loop power.
1For ambient temperature of 23 °C.
2Time from when the SETPOINT is reached to when the unit is with in stability specification.
3The temperature range may be limited by the reference probe connected to the readout.
The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterisation errors.
4Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05 Ωto the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
5The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MH.

Download

Fluke 9190A Datasheet

PDF Specifications of the Fluke 9190A Ultra-Cool Field Metrology Wells.

Download (2.34M)

No customer reviews for the moment.

Write your review

Fluke 9190A-X-P Field Metrology Well With Process Electronics
COMPONENT
DESCRIPTION PART No.
9190-INSA Insert "A" 9190A, imperial miscellaneous holes. 4231692
9190-INSB Insert "B" 9190A, imperial comparison holes. 4231704
9190-INSC Insert "C" 9190A, 0.25 inch holes. 4231719
9190-INSD Insert "D" 9190A, metric comparison holes. 4231728
9190-INSE Insert "E" 9190A, metric miscellaneous holes with 0.25 inch hole. 4231737
9190-INSF Insert "F" 9190A, metric comparison miscellaneous holes with 0.25 inch hole. 4231743
9190-INSZ Insert “Z” 9190, blank. 4231781
9190-INSY1 Insert "Y1" 9190A, custom insert with blank insulator. 4231770
9190-INSY2 Insert "Y2" 9190A, custom insert with custom insulator. 4319562
9190-CASE Carrying case for 9190A. 4231689

16 other products in the same category:

arrow_upward