- Hamilton01698 829711
- Leeds0113 235 1356
- Belfast02890 830302
Fluke P3211-PSI Single Piston Water Deadweight Tester
Includes: Fluke P3211-PSI accredited calibration report, weight mass details, M14 x 1.5, M20 X 1.5, 1/8 inch, 1/4 inch, 3/8 inch and 1/2 inch NPT and BSP female adaptors, operating fluid (where applicable) and spare seals.
Hydraulic Deadweight Tester - Water Operated – Single PCU - Pressure Range 10 to 500 psi.
These robust instruments are highly accurate, quick and easy to use. Units feature a built-in priming pump for large volume applications, piston flotation indicators and a high quality screw press for fine pressure control.
The P3200 Series is the culmination of over 50 years of experience in the production and design of high performance pressure standards. With features designed to improve accuracy and performance, increase reliability and simplify operation these deadweight testers can be used to calibrate virtually any pressure sensing device, including transducers, transmitters, gauges or pressure switches.
Each instrument is supplied with a detachable lid which makes it neat, compact and easily portable. The weight masses are stored in a high-quality case with a self locking mechanism to protect them during transit.
NB: Gravity varies significantly with geographical location and this variation has a direct effect on the force of the weights and the accuracy of the deadweight tester. Each instrument can be calibrated to local gravity at no extra cost. If unspecified, instruments will be calibrated to Standard Gravity at 980.665 cm/s2.
FEATURES
Distilled water operated: 10 to 500 psi.
0.015% of reading accuracy standard (0.008% optional).
Single piston format.
Built-in hand pumps standard.
Mounted spirit level with adjustable feet.
Units can be trimmed to local gravity FOC.
Distilled water operated | 10 to 500 psi. |
Accuracy | ±0.015% of reading (±0.008% optional). |
*Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. | |
Materials of Construction | |
Standard weight material | Series 3 non-magnetic austenitic stainless steel. |
Weight density | 7.8 g/cm³. |
Optional fractional weights | Solution heat treated aluminum. |
Weight density | 2.7 g/cm³. |
Piston material | Tungsten carbide with nickel binder Density- 15.0 g/cm³. |
Cylinder material | Hardened martensitic steel Tungsten carbide. |
Thermal coefficients of expansion | Oil piston/cylinder 16.5 ppm/°C. |
General Specifications | |
Test port adaptors | 1/8, 1/4, 3/8 and ½ NPT or BSP. |
Instrument weight | 36 lbs (16 kg). |
Instrument size | 17.5 W x 12 D x 8.5 H. |
Mass set weight (typical) | 80 lbs (36 kg). |
Reservoir volume | 150 cc. |
Screw press displacement | 5.5 cc. |
Pump displacement | 4.7 cc per stroke. |
O-ring seal materials | Buna N as standard, Viton and EPDM available. |
Weight Increments | |
Minimum Standard Weight Increments | Up to 500 psi (35 bar): 1 psi (0.1 bar) 500 to 5000 psi (35 to 350 bar): 10 psi (1 bar) 5000 to 20,000 psi (350 to 1400 bar): 20 psi (2 bar) Model P3112: 1 psi (0.1 bar). |
Optional Fractional Weights | Up to 500 psi (35 bar): 0.1 psi (0.01 bar) 500 to 5000 psi (35 to 350 bar): 1 psi (0.1 bar) 5000 to 20,000 psi (350 to 1400 bar): 2 psi (0.2 bar). |
Operating Fluid | |
Mineral oil | Shell spindle oil 22, our reference 55-655, viscosity 40 cs@ 20°C. |
Fluke P3211-PSI Warranty | One year. |
No customer reviews for the moment.
COMPONENT |
DESCRIPTION | PART No. | ||
P5551 | Pointer Remover/Punch. Designed to quickly remove and consistently refit the pointer of a pressure gauge. | 3875928 | ||
P5542 | Adaptor Set - NPT (1/8, 1/4, 3/8 and 1/2 NPT). | 3875943 | ||
P5521-70M | Liquid-to-liquid separators, 70 MPa (10k psi). | 4185715 | ||
P5522-50M | Liquid-to-liquid separators, 50 MPa (7000 psi), EP Seals. | 4185726 | ||
P5523-20M | liquid-to-gas Interface. 20 MPa (3000 psi). | 4335377 | ||
P5543 | Angle adaptor. | 3875955 |