loading spinner
Seaward Sentinel 501 Electrical Safety Tester
  • Seaward Sentinel 501 Electrical Safety Tester
  • Seaward Sentinel 501 Electrical Safety Tester
  • Seaward Sentinel 501 Electrical Safety Tester

Seaward Sentinel 501 Electrical Safety Tester

Seaward authorised distributor and service centre.

Includes: Clare Sentinel 501 User Manual, test lead set (hv cables x 2), interlock key, remote terminal plug, power lead, Test Lead Set (Earth Bond Cables x 2).

489A921

The Clare Sentinel 501 electrical safety tester (manufactured by Seaward) enables manufacturers to check their products in-house to ensure compliance with the relevant international technical conformance standards. Significant time and cost saving benefits are provided by the Sentinel 501 bench-top Hipot/flash tester from Clare, which deliver improved type testing during product development or enhanced compliance testing in a laboratory environment.

A remote start terminal and signal I/O port allows for remote starting, safety interlocking and PLC control while an enhanced safety feature will automatically cut off the output voltage within 150µs where normal output voltages are detected or when the upper or lower current trip limits are reached while the device is under test.

The Sentinel series is ideal for R&D facilities and compliance laboratories, while design, quality assurance, and manufacturing and production engineers will also benefit from the easy-to-use safety test features.

SENTINEL SERIES KEY FEATURES

500 VA capability (Sentinel 500 series).
Earth bond / insulation resistance (Sentinel 501).
Sweep function for test analysis.
Flexible / high resolution test parameters .
Manual / automatic test modes.
Remote and PLC control.
Fast cut off and automatic discharge.

Seaward Sentinel Series Comparison Chart
Features 200 500 501
Earth/ground Bond
AC dielectric strength - 5 kV
DC dielectric strength - 6 kV
DC insulation resistance
200 VA (40 mA)
500 VA (100 mA)
ARC detection
Seaward Sentinel 501 Specifications
AC Dielectric Strength
Programmable Voltage Range 0.100 kV - 5.000 kV (2V/Step Resolution).
Frequency Independent of Supply 50 or 60 Hz Selectable.
Output Voltage Accuracy ±1% + 5 Volts (No Load).
Voltage Display Resolution 0.001 kV.
Current Measurement 0.001 mA - 100.0 mA.
Range and Accuracy ±1.5% + 30 Counts when trip limit <1.11mA
±1.5% ± 3 Counts when trip limit >1.11mA.
Current Display Resolution 0.001 mA / 0.001 mA / 0.1 mA.
Selectable Range of Pass/Fail Levels 0.001 mA - 100.00 mA.
Maximum Rated Current 100 mA (0.5 kV ≤ V ≤ 5 kV)
10 mA (0.1 kV ≤ V ≤ 0.5 kV).
Arc Detection Yes.
Ramp time 0.1 - 999.9s.
Test Time Off / 0.5s - 999.9s.
DC Dielectric Strength
Programmable Voltage Range 0.100 kV - 6.000 kV
(2V/Step Resolution).
Frequency Independent 50 or 60 Hz Selectable of Supply 50 or 60 Hz Selectable.
Output Voltage Accuracy ±1% + 5 Volts (No Load).
Voltage Display Resolution 0.001 kV.
Current Measurement 0.001 mA - 20.0 mA.
Range and Accuracy ±1.5% + 30 Counts when trip limit <1.11 mA
±1.5% + 3 Counts when trip limit >1.11 mA.
Current Display Resolution 0.001 mA / 0.001 mA / 0.1 mA.
Selectable Range of Pass/Fail Levels 0.001 mA - 20.00 mA.
Maximum Rated Current 20 mA (0.5 kV ≤ V ≤ 5 kV)
2 mA (0.1 kV ≤ V ≤ 0.5 kV).
Arc Detection Yes.
Ramp time 0.1 - 999.9s.
Test Time Off / 0.5s - 999.9s.
Earth Bond
Programmable Test Current 03.00 A - 32.00 A ac.
Test Current Resolution 0.01 A.
Test Current Accuracy 3 A - 8 A: (1% of rdg+0.2 A)
8 A - 32 A: (1% of rdg+0.05 A).
Test Voltage 6 V AC max (open circuit).
Test Voltage Frequency 50 Hz/60 Hz selectable.
Resistance Measurement Range 10mΩ - 650.0 mΩ.
Resistance Measurement Resolution 0.1 mΩ.
Resistance Measurement Accuracy (1% of rdg + 2 mΩ).
Test Time 0.5 - 999.9s.
Arc Detection Yes.
Insulation Resistance
Output Voltage 50 V - 1000 V.
Output-Voltage Resolution 50 V / Step.
Output-Voltage Accuracy ±(1% of setting +5 V)[no load].
Resistance Measurement Range 0.001 GΩ -50.00 GΩ.
Test Voltage
50 V - 450 V 0.001 - 0.050 GΩ
0.001 - 0.500 GΩ.
500 V - 1000 V 0.051 - 2.000 GΩ
0.501 - 9.999 GΩ
10.00 - 50.00 GΩ.
Selectable Range of Pass/Fail Levels 0.001 mA - 20.00 mA.
Output Impedance Yes.
RAMP (Ramp-Up Time) 600 kΩ.
TIMER (Test Time) 1s~999.9s.
General Specifications
Dimensions 330W x 148H x 587D mm.
Weight Approx 27 kg.
Sentinel 501 warranty 1 year (subject to product registration with Seaward).

Download

Seaward Sentinel Series

PDF Specifications of the Seaward Sentinel Electrical Safety Testers.

Download (307.97k)

No customer reviews for the moment.